The improvement of tribological parameters of nano structural multilayer coatings for unipolar modular joint applications

Authors

DOI:

https://doi.org/10.15674/0030-59872014128-33

Keywords:

hemiarthroplasty, implant materials, nano structural multilayer coatings, tribological performance, biocompatibility

Abstract

Femoral neck fractures are a common insufficiency fracture in the elderly. The application of ceramic coatings on metallic materials represents an effective way to formation of advanced surfaces of unipolar and bipolar modular prosthesis for joint arthroplasty. The enhancement of mechanical properties such as hardness parameters up to 14 GPa, adhesion strength up to 50 N and low friction coefficient of multilayer coatings, with strong tendency to decrease in fluids (0.07), effective Young's modulus, toughness, elastic recovery and wear resistance of the coatings is very important for tribological performance of the next generation of ceramic coated modular prosthesis. The joint femoral head surfaces with nano structural multilayer coatings demonstrate improved biocompatibility due to wettability increasing and fluid friction characteristics advancing.

References

  1. Ravikumar K. J. Internal fixation versus hemiarthroplasty versus total hip arthroplasty for displaced subcapital fractures of femur 13 year results of a prospective randomised study / K. J. Ravikumar, G. Marsh // Injury. — 2000. — Vol. 3, № 10. — P. 793–797.
  2. Frihagen F. Hemiarthroplasty or internal fixation for intracapsular displaced femoral neck fractures:randomized controlled trial / F. Frihagen, L. Nordsletten, J. E. Madsen // BMJ. — 2007. — Vol. 335 (7632). — P. 1251–1254.
  3. The human cost of fracture / J. A. Pasco, K. M. Sanders, F. M. Hoekstra [et al.] // Osteoporos Int. — 2005. — Vol. 16 (12). — P. 2046–2052.
  4. Unipolar or Bipolar Prosthesis for displaced intracapsular hip fracture in octogenarians: a randomized prospective study / S. J. Calder, G. H. Anderson, C. Jagger [et al.] // J. Bone Joint Surg. — 1996. — Vol. 78 (3). — P. 391–394.
  5. Unipolar versus bipolar hemiarthroplasty for the treatment of femoral neck fractures in the elderly / C. N. Cornell, D. Levine, J. O’Doherty, J. Lyden // Clin. Orthop. Relat. Res. —1998. — P. 67–71.
  6. Unipolar versus bipolar hemiarthroplasty: functional outcome after femoral neck fracture at a minimum of thirty six months of follow up / B. C. Ong, S. G. Maurer, G. B. Aharonoff [et al.] // J. Orthop. Trauma. — 2002. — Vol. 16 (5). — P. 317–322.
  7. Unipolar or bipolar hemiarthroplasty for femoral neck fractures in the elderly? / F. J. Raia, C. B. Chapman, M. F. Herrera [et al.] // Clin. Orthop. Relat. Res. — 2003. — Vol. 414. — P. 259–265.
  8. Varley J. Stability of hip arthroplasties / J. Varley, M. J. Parker // Int. Orthop. — 2004. — Vol. 28, № 5. — P. 274–277.
  9. Bohm E. R. Employment status and personal characteristics in patients awaiting hip-replacement surgery / E. R. Bohm // Can. J. Surg. — 2009. — Vol. 52, № 2. — P. 142–146.
  10. Morscher E. W. Failures and successes in total hip replacement Review / E. W. Morscher // Scand. J. Surg. — 2003. — Vol. 92. — P. 113–120.
  11. Patient characteristics affecting the prognosis of total hip and knee joint arthroplasty: a systematic review / P. L. Santaguida, G. A. Hawker, P. L. Hudak [et al.] // Can. J. Surg. — 2008. — Vol. 51, № 6. — P. 428–436.
  12. Petsatodis G. E. Primary cementless total hip arthroplasty with an alumina ceramic-on-ceramic bearing: results after a minimum of twenty years of follow-up / G. E. Petsatodis // J. Bone Joint Surg. — 2010. — Vol. 92-A, № 3. — P. 639–644.
  13. Lymphocyte proliferation responses in patients with pseudotumors following metal-on metal hip resurfacing arthroplasty / Y. M. Kwon, P. Thomas, B. Summer [et al.] // J. Orthop. Res. — 2010. — Vol. 28, № 4. — P. 444–450.
  14. Patient activity after total hip arthroplasty declines with advancing age / S. Kinkel, N. Wollmerstedt, J. A. Kleinhans [et al.] // Clinic. Orthop. — 2009. — Vol. 467, № 8. — P. 2053–2058.
  15. Campbell P. Histological features of pseudotumor-like tissues from metal-on-metal hips / P. Campbell // Clinic. Orthop. — 2010. — Vol. 468, № 9. — P. 2321–2327.
  16. Comparison of the cytotoxicity of clinically relevant cobalt-chromium and alumina ceramic wear particles in vitro / M. A. Germain, A. Hatton, S. Williams [et al.] // Biomaterials. — 2003. — Vol. 24, № 3. — P. 469–479.
  17. Ceramic acetabular linear fracture in total hip arthroplasty with a ceramic sandwich cup / M. Hasegawa, A. Sudo, H. Hirata, A. Uchida // J. Arthroplasty. — 2003. — Vol. 18, № 5. — P. 658–661.
  18. Metal wear particle characterization from metal-on-metal hip replacements: Transmission electron microscopy study of periprosthetic tissues and isolated particles / P. F. Doom, P. A. Campbell, J. Worrall // J. Biomed. Mater. Res. — 1998. — Vol. 42. — P. 103–111.
  19. Merritt K. Distribution of cobalt chromium wear and corrosion products and biologic reactions / K. Merritt, S. A. Brown // Clinic. Orthop. — 1996. — Vol. 329. — P. 233–243.
  20. Dissemenation of wear particles to the liver, spleen and abdominal lymph nodes of patients with hip or knee replacement / R. M. Urban, J. J. Jacobs, M. J. Tomlinson [et al.] // J. Bone Joint Surg. — 2000. — Vol. 82-A. — P. 457–476.
  21. Nanocomposite oxide and nitride hard coating produced by pulse magnetron sputtering / H. Klostermann, B. Bufcher, F. Fietzke [et al.] // Surface&Coating Tech. — 2005. — Vol. 200. — P. 760–764.
  22. Direct current magnetron sputtering deposition of nanocomposite alumina-zirconia thin films / D. H. Trinh, T. Kubart, T. Nyberg [et al.] // Thin Solid Films. — 2008. — Vol. 516. — P 352–8358.
  23. The corrosion properties of zirconium and titanium load-bearing implant materials with protective oxide coatings / A. Zykova, V. Safonov, J. Smolik [et al.]: Proceeding of 13th International Conference on Plasma Surface Engineering, 2012. — Р. 10–14.
  24. Cell adhesion on biomaterial surfaces with nanocomposite oxide coatings / O. Vyrva, A. Zykova, V. Safonov [et al.] // Orthopaedics, traumatology and prosthetics. — 2011. — Vol. 2. — P. 84–87.
  25. Nano scale biomaterial interface modifications for advanced tissue engineering applications / V. Safonov, A. Zykova, J. Smolik [et al.] // J. Phys. Conf. Ser. — 2012. — Vol. 356. — Р. 012–046.
  26. Kusano Y. Methods of data analysis for the micro-scale abrasion test on coated substrates / Y. Kusano, K. Van Acker, I. M. Hutchings // Surf. Coat. Technol. — 2004. — Vol. 183, № 2–3. — P. 312–327.
  27. Camara Cozza R. A study on friction coefficient and wear coefficient of coated systems submitted to micro-scale abrasion tests / R. Camara Cozza // Surf. Coat. Technol. — 2013. — Vol. 215. — P. 224–233.
  28. Leyland A. Design criteria for wear resistant nano structured and glassy-metal coatings / A. Leyland, A. Mathews // Surf. Coat Technol. — 2004. — Vol. 177, 178. — P. 317–321.
  29. Musil J. Two-phase single layer Al-O-N nanocomposite films with enhanced resistance to cracking / J. Musil, J. Sklenka, R. Cerstvy // Surf. Coat. Technol. — 2012. — Vol. 206. — P. 4230–4234.
  30. Musil J. Transparent Zr-Al-O nanocomposite coatings with enhanced resistance to cracking / J. Musil, J. Sklenka, R. Cerstvy // Surf. Coat. Technol. — 2912. — Vol. 206. — P. 2105–2109.

Downloads

How to Cite

Zykova, A., Safonov, V., Lukyanchenko, V., Samoilenko, O., Korzh, M., Yakovin, S., Kolesnikov, D., Goncharov, I., & Donkov, N. (2014). The improvement of tribological parameters of nano structural multilayer coatings for unipolar modular joint applications. ORTHOPAEDICS TRAUMATOLOGY and PROSTHETICS, (1), 28–33. https://doi.org/10.15674/0030-59872014128-33

Issue

Section

ORIGINAL ARTICLES